Portable Ultrasonic Flowmeter Installation Precautions

Jan 11, 2025

Leave a message

 

1. The liquid must be filled with the pipe, find out and accurately input the pipe diameter, material, wall thickness, lining material and lining thickness of the pipe related parameters, the pipe wall to remove the insulation layer, need to be polished;


2. When installing the transducer according to the installation distance calculated by the flow meter, and comprehensively referring to the signal strength of the flow meter, measured sound velocity value, signal quality, installation distance inspection and other parameters, the transducer installation position is adjusted to the best position. Note that the two probes should be installed in the relative diameter of the pipe, the pipe is thicker to be accurately installed, carefully adjusted (can use paper to find the exact position), if the flow meter measured sound velocity value and the actual sound velocity value difference of more than ±6 m/s, the flow meter measured flow value and the actual flow value will also be different;


3. When the transducer is installed, the front and back straight pipe section should meet the requirements of at least 5D after D, avoid the top and bottom of the pipeline, try to install in the horizontal position of the pipeline, or within the range of 45° with the horizontal diameter, pay attention to avoid the weld. A layer of acoustic coupler applied between the probe and the pipe or workpiece surface;


4. The flow meter in different pipe diameter verification, calibration, will get different correction coefficients, even if the same host using the same type of different transducers, will get different correction coefficients. Therefore, users who use a single ultrasonic flowmeter to measure different pipe diameters should be regularly inspected in accordance with the requirements of the national metrological verification regulations.

Send Inquiry