Problems of pipeline ultrasonic flowmeter

May 16, 2024

Leave a message

 

Pipe type ultrasonic flowmeter is also called pipe segment type ultrasonic flowmeter, is currently on the market for flow measurement is a very common measuring tool, especially suitable for flow balance test and flow monitoring: drinking water, river water, seawater, cooling water, hot water, industrial sewage, lubricating oil, diesel, fuel oil chemical liquids.

 

The shortcomings and treatment of pipeline ultrasonic flowmeter:

 

1. No signal

 

The pipeline type ultrasonic flowmeter has no signal, generally because the transducer does not have a useful connection with the host. The join will need to be reconnected while processing is taking place. The loss of the signal may also be due to power supply pollution, and it is necessary to purify the power supply in order to deal with the problem. In addition to the above reasons, the change in the orientation of the transducer will also cause the signal to be affected, and its orientation needs to be adjusted;

 

2. Instantaneous flow greater than the practice value

 

The reason for this defect is generally because the transducer acoustic wedge has a defect, and the transducer needs to be replaced;

 

3. Instantaneous flow has a large fluctuation

 

The reason for this shortcoming is mainly because the fluid is disturbed, causing its flow state to appear unstable, or the pipe type ultrasonic flowmeter host does not set the damping coefficient to the specified scale, making its value small. When dealing with such defects, the position of the transducer can be adjusted. Secondly, the damping coefficient needs to be increased;

 

4. Instantaneous and cumulative flow is different

 

The main reason for such shortcomings is generally that the pipe type ultrasonic flowmeter host presents problems, and the shortcomings need to be dealt with by replacing the host.

Send Inquiry